ROHO SelectAir MA mattress User Manual

Page 24

Advertising
background image



Note: To assist in deflating the mattress, disconnect the hose at the

mattress or control unit.

Then press the "Upright On/Off" button a second time (Ensure the upright

light extinguishes).

Note: Do not leave the control unit in the Upright Mode when the

patient is lying flat or at a slight angle.

Note: Ensure the patient has at least one (1) inch (3 cm) of clearance

between the bed frame and their body. If not, manually increase or

decrease system pressure to compensate.

Firm Mattress Mode

To aid in the placement and maneuvering of the patient, the control unit is

equipped with a Firm Mattress Mode. The Firm Mattress Mode will fully inflate

the mattress making it firm for easy positioning. The Firm Mattress Mode will

automatically terminate when the time displayed on the keypad has expired.

Press the "Firm" button on the keypad. The display will indicate the

remaining time (fig. 15).

To end the cycle press the Firm Mattress “Off” button.

Note: The SelectAir control unit is equipped with an internal timer which will

turn the Firm Mattress Mode off after 0 minutes. If the Firm Mattress

Mode has timed out and ended the firm mattress cycle, DO NOT press

the Firm "Off” key as this may cause the system to "lock up".

Note: Pressing the Firm "Off" key multiple times or holding down the button

may cause the system to “lock up”.

Note: To ensure patient comfort, the control unit slowly lowers the pressure of

the system. This may give the appearance that the firm function is still

active. If the timer lights are extinguished, the control unit is in its

normal mode.

.

.

Section 0

Select

Air

Figure 6

Select

Air,

Select

Air Max,

HybridSelect

Models

Flashing

“Green” Light

(Power to

Control Unit)

CPR/Emergency Deflate

Important: In case of a CPR Emergency, turn power OFF by pressing the

CPR button on the keypad.