HealFloat feature (HybridSelect only)

Refer to the ROHO HIGH PROFILE cushionʼs operating manual for specific cleaning

instructions.

Control Unit and Hose (between patients)

1. Unplug the control unit from the wall outlet before proceeding.
2. Wipe down entire control unit and all hoses with a disinfectant solution. MOXI

recommends using LpHse

®

diluted 1:256.

3. Allow solution to remain on system for 10 minutes.
4. Wipe down entire control unit and dry with disposable cleaning cloth or sponge.
5. Only use a mild detergent (dishwashing) and warm water to clean the filter. Rinse

the filter with cool water and thoroughly air dry the filter before replacing.

All Control Units

DANGER:

Disconnect control unit from main power supply before cleaning.

Failure to do so could result in electric shock.

Note:

•Do not immerse the control unit in liquids.

•Do not heat or steam autoclave the control unit.

Note:

Do not saturate the surface of the control unit with liquids. Wipe

off any dust. If necessary, clean the housing exterior with a

disinfectant solution or mild detergent and a damp cloth. Wipe

dry.

DANGER:

Make sure the control unit is completely dry before attempting to

reconnect or operate it.

Section 8

Cleaning and Disinfecting