9

• To lock the control panel, press the “Lock/Unlock” button. If no patient activity

is detected for 60 seconds, the control unit will automatically lock the system.

Auto Comfort

The BariSelect control unit is equipped with the Auto

Comfort mode. When selected, the control unit will

proceed through one or several (depending on the

patient's body type) cycles and will then calculate the

approximate proper setting for the patient. If the setting

selected by the control unit is not acceptable, then a

second run of the Auto Comfort cycle may be necessary.

If manual setup is desired see Manual Setup, Section 6.

Turn the control unit on and allow the mattress to fully inflate before making

any adjustments or placing the patient on the mattress.

Place the patient on the mattress, centered and in their habitual resting

position, i.e. on their back, on their side, etc.

Note: The bed rails should be locked in their highest position before

proceeding.

Deactivate the Firm Mode before starting Auto Comfort.

Press the Auto Comfort "Start" button (Fig. 29).

Note: The patient must remain still until the cycle is complete.

Note: Do not press any other buttons until the Auto Comfort cycle completes.

Inspect the patient to ensure they have sunk into the mattress about 40-50%

and not touching the frame at any point. If they have not, adjust the Comfort

Level using the Comfort Level "Softer" and "Firmer" buttons until the proper

results are achieved.

Note: Because weight distribution varies from patient to patient, some

patients may require an increase or decrease in the Comfort Level.