By LogBox Button: Press the Start_Now button from the Monitoring screen,

the Palmtop should be pointed towards the logger.

By Digital Input: Readings are performed while the digital input is enabled

(closed / 1) and interrupted when the digital input is disabled (open / 0).

Stop:

Defines logging stop mode: Options are:

Full memory: Loggings can be stored up to the logger full memory capacity

is reached.

Wrap around: Logging never stops. The LogBox will keep on recording the

readings and when the memory is full it will overwrite the oldest record in a

circular or wrap around manner.

After loggings: The logging will stop after the number of readings here

defined.

By date/time: Logging is stopped on user-defined day and time.

Interval:

Defines the interval between readings: hour, minute and second. For mean,

maximum and minimum values, the shortest interval between loggings is 10

seconds.

Channel 1: Opens the Input 1 Settings screen.

Tag: Defines a name for Channel 1.

Input: Informs the input type used in Channel 1:

Unit: Defines the unit of the variable. Options are set up for flow measurement,

however, other units can be defined by typing the unit in the text box.

Input Factor: Defines the amount of input per pulse. It is set up to measure

liquid flow (liters per pulse.) It may be a value between 0.000001 and 16383.

Scale Factor: Adjusts the readings such as to represent a particular flow

measurement unit. This factor can be automatically calculated by tapping on

the button Calc. It may be a value between 0.000001 and 65535.

Alarms: Enables an alarm that is triggered according to user-defined

parameters.

Alarm settings screen

Cancel and OK buttons cancel and save configurations defined in Channel 2 screen.