6�- TECHNICAL FEATURES

AC Power supply

Rated voltage

100���520V

~

Frequency

50���60Hz

Consumption

4���10VA

Installation category

CATIII300V

Voltage measurement circuit

Rated voltage (Un)

230V

P-N

,400V

P-P

Voltage measurement margin

20to300V

P-N

,35to520V

P-P

Frequency measurement margin

45���65Hz

Input impedance

660kΩ

Minimum measurement voltage (Vstart)

20V

P-N

,35V

P-P

Installation category

CATIII300V

Current measurement circuit

Nominal current (In)

���/5Aor���/1A

Current measurement margin

1to120%In

Minimum measurement current (Istart)

50mA

Installation category

CATIII300V

Leakage current measurement circuit

Bymeansofanearthleakagetransformerwitha

ratioof500turns

Nominal current of the secondary

3mA

Current measurement margin

10mAto1�5A

Minimum measurement current (Istart)

10mA

Measurement accuracy UNE-EN 61557-12

Voltage measurement

0�5%±1digit

Current measurement

0�5%±1digit

Active power measurement

0�5%±2digits

Reactive power measurement

1%±2digits

Active energy measurement

Class1

Reactive energy measurement

Class2

Digital outputs

Model

Computer SMART III FAST 6 Computer SMART III FAST 12

Quantity

6

2

6

2

Type

OPTOMOS

NPN

OPTOMOS

NPN

Maximum voltage

60VDC

24VDC

60VDC

24VDC

Maximum current

0�2A

50mA

0�2A

50mA

Relay outputs

Quantity

2(fan,alarm)

Maximum voltage, open contacts

1kV

Maximum current

1A

Maximum switching power

2500VA

Electrical life

30x10

3

cycles

Mechanical working life

5x10

6

cycles