Bronkhorst RS232 interface User Manual

Page 12

Advertising
background image

BRONKHORST

®

 

Page 12   

 

 

 

 

RS232 interface   

 

 

 

9.17.027 

3.3.3 RS232 ASCII protocol 

 
An ASCII protocol is used on the existing RS232‐HOST. To be compatible with existing driver software the ASCII 
protocol is available. 
 
A basic data link message is coded in ASCII as follows: 
 

 

length 

node

data

 

: 

len1 

len2

node1

node2

data1

data2

CR 

 

Byte 

Explanation 

Initial character (semicolon)

Len1, len2 

Length of message including the node address in bytes, so (len1, len2) is 
the basic message length +1. 

node1, node2

node address of destination (PC to HOST)
node address of source (HOST to PC) 

data1, data2 

message field

CR 

termination character

 
All bytes (except the initial and termination character) are converted from 1 binary byte to 2 hexadecimal bytes in 
ASCII representation. 
 

Example: binary data byte 0x2A ‐‐> hexadecimal ASCII characters 0x32, 0x41. 

 
A special message type is used to pass error messages from the HOST to the PC. Its structure is as follows: 
 

 

0x01

error

: 

0x30

0x31

error1

error2

CR

Byte 

Explanation 

initial character

0x30, 0x31 

length of the message (1 byte)

error 

error code, two digit HEX number

CR 

termination character


The error code can have the following values: 
 

Value 

Meaning 

propar protocol error

propar protocol error (or CRC error)

destination node address rejected

response message timeout

1,2,8 

general error 

 

Advertising