Rs232/flow‐bus – Bronkhorst RS232 interface User Manual

Page 19

Advertising
background image

BRONKHORST

®

 

Page 19   

 

 

 

 

RS232 interface   

 

 

 

9.17.027 

3.9

I

NITIALISING 

RS232/FLOW‐BUS

 INTERFACE

 

 
If  you  use  a  RS232/FLOW‐BUS  interface  for  communication  (without  the  micro‐switch  and  2  LED’s),  note  that  this 
module is not part of the (FLOW‐BUS) token‐ring network, directly at power‐up. This means that it is always necessary 
to re‐initialise  the module when  power has been interrupted! This is not the case when using an RS232/FLOW‐BUS 
interface with micro‐switch, red LED, green LED and RJ45 connector for FLOW‐BUS. By means of the switch you may 
force  the  interface  to  find  a  free  address  on  the  FLOW‐BUS  once.  You  may  skip  the  initialisation  and  start  directly 
sending messages. 
 
Also when using digital (Multibus) instruments with RS232 directly on the instrument it is not needed to initialise (give 
a  free  node‐address  to)  the  instrument  on  the  FLOW‐BUS  because  instrument  is  not  physically  connected  to  the 
FLOW‐BUS,  but  only  uses  the  same  protocol.  You  may  start  directly  sending  your  messages  to  the  instrument  on 
either  the  node‐address  of  the  instrument  in  its  memory  e.g.  node  3  (selective  response)  or  to  node‐address  128 
(always response).  
 
At power‐up situation you can communicate with the RS232 interface only at the RS232 side via node 0. To get part of 
the FLOW‐BUS you have to send an init command, send the network parameters PNA, SNA, NNA, LNA and BM and 
send a reset command. From this moment the interface is part of the FLOW‐BUS. Ensure the module gets a free and 
unique address on the bus, 2 modules on the same address will cause communication problems. When you are sure 
that  there  are  no  more  interfaces  in  the  system,  simply  force  the  RS232/FLOW‐BUS  interface  to  address  1.  This 
address is reserved for an interface. PC‐support software (FLOWB32.DLL) will search for a free address on which the 
interface will be installed. 
 
Follow the steps below to realise correct initialisation for this interface via RS232: 
 

Initialisation RS232 interface (needed for FLOW‐BUS/RS232 interfaces without switch and LED’s only)

Send 

Response 

Comment

:050001000A49\r\n 

 

Init instruction for node 0 process 0.

 

:04000000XX\r\n 

No error.

:050001000101\r\n 

 

PNA = Primary Node Address = 1

 

:04000000XX\r\n 

No error.

:05000100027F\r\n 

 

SNA = Secondary Node Address = 127

 

:04000000XX\r\n 

No error.

:050001000302\r\n 

 

NNA = Next Node  Address = 2

 

:04000000XX\r\n 

No error.

:050001000420\r\n 

 

LNA = Last Node Address = 32 (depends on system size) 

 

:04000000XX\r\n 

No error.

:050001000502\r\n 

 

BM = Bus Management = 67 
67 = everything automatically (auto arbitration + gap skipping) 
 3  = auto arbitration 

2  = always bus master 
1  = temporary  
In older systems: when no R/C‐modules in system make BM = 2, 
when R/C‐modules in system (already bus masters present) than 
make BM = 1; otherwise make = 67 

 

:04000000XX\r\n 

No error.

:050001000A52\r\n 

 

Reset instruction for module; from this moment on module will be 
active on FLOW‐BUS at node address  = PNA 

 

:04000000XX\r\n 

No error.

 

 

Sometimes it could be necessary to repeat the first instruction. Wait approx. 2 seconds before 

sending the next command. 

XX means: don’t care 

 
 
 