10 

Parameter index 16 (return)

Parameter number 31

Capacity unit 

21 

01 

00000001  P

01 

Process 1

22 

7F 

01111111 

60 

Parameter type ‘string’

P

1F 

Parameter number (FBnr.) 31 – Capacity unit

23 

07 

00001110   

07 

String length 7

24 

71 

01110001 

C

00 

Parameter not chained (return)

Parameter index 17 (return)

Process 1 

Parameter number 17 

Fluid name 

60 

Parameter type ‘string’ (return)

11 

Parameter index 17 (return)

25 

01 

00000001  P

01 

Process 1

26 

71 

01110001 

60 

Parameter type ‘string’

P

11 

Parameter number (FBnr.) 17 – Fluid name

27 

0A 

 

 

 

String length 10

29 

\n   

 

 

Line Feed

 

28 

\r 

 

 

 

Carriage Return

End

Total String 

:1A0304F1EC7163006D71660001AE0120CF014DF0017F077101710A\r\n 

 

Answer by node 3

Nr 

Byte  Layout 

Description 

Description per block

 

 

Start 

37 

 

Number of bytes which do follow: 65 bytes

03 

 

Node 3 

02 

 

Command write

F1 

11110001 

80 

Process chained

Process 113 

71 

Process 113 (receiving process)

EC 

11101100 

80 

Parameter chained

Parameter index 12 

Serial number 

60 

Parameter type ‘string’

0C  Parameter index 12

00 

 

Length of the answer 10 Bytes

7‐26 

 

 

4D 36 32 31 32 33 34 35 41 00
Parameter value converted from hex to ASCII : M6212345A 

27 

6D 

01101101 

00  Process not chained

Parameter index 13 

USERTAG 

60  Parameter type ‘string’

0D  Parameter index 13

28 

00 

 

String length 00, length not defined

29‐36 

 

 

55 53 45 52 54 41 47 00
Parameter value converted from hex to ASCII, the values do 
read : USERTAG 

37 

01 

00000001 

00  Process not chained

Process 1 

01  Process 1 (receiving process)

38 

AE 

10101110 

80  Parameter chained

Parameter index 14 

Measure 

20  Parameter type ’integer’

0E  Parameter index 14

39 

1C 

 

Parameter value is: 1CD8 (hex)
Measure Value is:  7384 (dec) 

 

40 

D8 

41 

CF 

11001111 

80  Parameter chained

Parameter index 15 

Capacity 

40  Parameter type ‘float’

0F 

Parameter index 15

42 

3F 

 

3F 80 00 00 
Parameter Value in IEEE‐floating point notation, 32‐bit single 
precision  
Parameter value converted from float to decimal, the values 
reads : 1.0 

43 

80 

44 

00 

45 

00 