Bronkhorst RS232 interface User Manual

Page 26

Advertising
background image

BRONKHORST

®

 

Page 26   

 

 

 

 

RS232 interface   

 

 

 

9.17.027 

20  Parameter type ‘integer’

measure

01  Parameter index 1

3E 

 

Value 3E80h = 16000 = 50% 

80 

\r 

Carriage Return 

End 

\n 

Line Feed 

Total String 

:06030201213E80\r\n

 

3.10.6 Request counter value 

 

Request Counter value from node 3, process 104, type float

Nr  Byte  Layout 

Description 

Description per block

 

Start 

06 

Length 6 

03 

Node 3 

04 

Command read 

68 

01101000 

C

00 

Process not chained (return)

Process 104 

P

68 

Process 104 (return)

41 

01000001 

C

00 

Parameter not chained (return)

Parameter index 1 

Process 104 

Parameter number 1 

Counter value 

40 

Parameter type ‘float’ (return)

01 

Parameter index 1 (return)

68 

01101000  P

68 

Process 104

41 

01000001 

40 

Parameter type ‘float’

P

01 

Parameter number (FBnr.) 1 (counter value)

\r 

 

 

 

Carriage Return

End 

\n 

 

 

 

Line Feed

Total String 

:06030468416841\r\n

 

Answer from node 3, process 104

Nr  Byte  Layout 

Description 

Description per block

0 :

 

Start 

08 

 

Length 8 

03 

 

Node 3 

02 

 

Command write 

68 

01101000 

C

00 

Process not chained 

Process 104 

P

68 

Process 104 (receiving process)

41 

01000001 

C

00 

Parameter not chained

Parameter index 1 

Counter value 

40 

Parameter type ‘float’

01 

Parameter index 1

45 

 

Parameter value ‘float’ = 5023.96 decimal 

9C 

FF 

AE 

10 

\r 

 

Carriage Return 

End 

11 

\n 

 

Line Feed 

Total String 

:0803026841459CFFAE\r\n

 
 