To Enter a New Setup and Calibration Number
Short Form Calibration

The Short Form Setup & Calibration procedure allows you to change the “SETUP” and “CAL”

numbers of the indicator. You may want to perform this procedure if the indicator is being

connected to different load cells or if the scale is inaccurate and you have some accurate

known weights to use for calibrating the scale.
Before continuing, first write down the current SETUP and CAL numbers of your EZ

indicator. These numbers are displayed during the Self Test.
To run the self test with the indicator already ON:

Press the [NET/GROSS] key and then press the [ON/RECORD] key to start the Self Test.

Press the [ON/RECORD] key to "pause" the Self Test while numbers are displayed.

Press the [ON/RECORD] key again to allow the self-test to complete normally.

.SETUP # _______________ CAL # ________________

Note:

Do not attempt to calibrate the scale if the indicator is not reading stable weights. The

calibration procedure will not fix instability, inconsistencies, or flashing "RANGE" messages.

Press and hold the [ZERO/RECHECK] key and press the [ON/RECORD] key for 3 seconds

to enter the short form calibration. The display will flash “SETUP” and then display the 6-

digit setup number with the right digit flashing. To modify the setup number:

1. Press the [Net/Gross] key several times to increment the digit to it proper value.
2. Press the [TARE] key to advance the blinking digit to the left.
3. Repeat steps 1 and 2 for each digit as required.
4. Press the [ON/RECORD] key to enter the new setup number and display the

calibration number.

5. Repeat steps 1 and 2 to modify the calibration number.
6. Press [ON/RECORD] to enter the new calibration number and the display will go

back to normal.

Verify the accuracy of the scale and the procedure is complete!