Installation

continued

Measuring Sensor ORG 12, ORG 22

continued

Sensor installed in a sample line,
e. g. water treatment, beer
filtering, permeate downstream
of reverse osmosis.

Installation of sensor when the
condensate contains suspended
solids. If the sensor is installed in
acc. with figs. 4 – 6, solids could
deposit on the cleaning device
and scratch the glass cylinder .

Sensor installed in a bypass of a
product line with upward flow.

Sensor installed below a product
line carrying a slight amount of
gas, e. g. sufficiently sized
condensate line (flashing)
downstream of steam traps.

*) Tap product line at the side
(angle of 45°) from the bottom to
prevent air and dirt from flowing
into the bypass.

Inlet Outlet

Cleaning device V-shaped
flow through sensor,
automatic deaeration

Cleaning device

Outlet

Inlet

H

500

Fig.2

Fig.3

Fig.4

Fig.5

*)

U