Commissioning procedure – Flowserve OR 52-5 User Manual

Page 26

Advertising
background image

26

Commissioning Procedure

continued

Adjusting Limit Alarm 1

1. Press , the yellow P-LED lights up.
2. Use

to change the LED display until LED Alarm 1

and LED ppm light up. The value saved last is shown.

3. Press , the yellow P-LED and the second digit of the

7-segment display are flashing.

4. Use

to change the digit, press to save the

setting and move to the third digit. The third digit is
now flashing.

5. Use

to change the digit, press to save the

setting. The P-LED and the 7-segment display are
illuminated. A limit value of max. 15 ppm can be adjusted.

6. Press , the equipment returns to normal operation, the actual value is indicated and

the LED ppm is illuminated.

Adjusting Limit Alarm 2

1. Press , the yellow P-LED lights up.
2. Use

to change the LED display until LED Alarm 2

and LED ppm light up. The value saved last is shown.

3. Press , the yellow P-LED and the second digit of the

7-segment display are flashing.

4. Use

to change the digit, press to save the

setting and move to the third digit. The third digit is now
flashing.

5. Use

to change the digit, press to save the

settings. The P-LED and the 7-segment display are
illuminated. A limit value of max. 15 ppm can be adjusted.

6. Press , the equipment returns to normal operation, the actual value is indicated and

the LED ppm is illuminated.

ALARM

1

2

ppm

sec

cal

0

OIL

Test

P

E

P

P

E

E

E

ALARM

1

2

ppm

sec

cal

0

OIL

Test

P

E

P

P

E

E

E