Indication of Error Code

In the event of a malfunction the yellow LED Malfunction
lights up, the relay output Malfunction opens and the
7-segment display indicates the error code.

The raising of an alarm is not effected by a malfunction.

Error Code Table

Error code

Possible reason

Remedy

E 01

Glow lamp in light emitter

contaminated

Replace glow lamp

Glass cylinder

contaminated

Clean or replace glass cylinder

Turbidity of fluid too high

Provide mechanical cleaning stage
ahead of equipment

Operating voltage too low

Check supply voltage

Cleaning device

not in its initial

position

Return cleaning device

to its

initial position

Ingress of solids

Provide mechanical cleaning stage
ahead of equipment