Maintenance, Error code – Flowserve OR 52-5 User Manual

Page 32

Advertising
background image

32

Maintenance

continued

Error Code Table

continued

Error code

Possible reason

Remedy

E 03

Glow lamp in light emitter
defective. A frequent occurrence
of this problem indicates heat
accumulation in light emitter
(despite fluid temp. > 60 °C
drying cartridge mounted)

Replace glow lamp. Remove
drying cartridge

and install

vent nipple

Cable connection to light
emitter

interrupted

Check connecting cables

Cleaning device

not in its

initial position

Return cleaning device

to its

initial position

Ingress of solids

Provide mechanical cleaning stage
ahead of equipment

Error code

Possible reason

Remedy

E 05

Basic turbidity > 10 ppm

Provide mechanical cleaning stage
ahead of equipment

Glass cylinder

scratched

(ingress of solids)

Replace glass cylinder

.

Install sensor acc. to installation
example fig. 3

Glass cylinder

contaminated

Clean or replace glass cylinder

Ingress of solids

Provide mechanical cleaning stage
ahead of equipment

Presence of gas or steam
bubbles (flashing)

Throttle outlet

Error code

Possible reason

Remedy

E 06

Light receiver

defective or

incorrectly connected.

Replace light receiver

or check

connecting cables

Measuring transducer defective

Replace measuring transducer

Q

Q

G

F

Q

H

H

U

U

U

U

D

D