Interpreting the optical image and diagnostic values

The diagnostics tab on the instruments webpage contains most of the tools for troubleshooting. The optical

image can be used for inspecting the condition of the optics. The diagnostic values provide information about

the measurements, conditions inside the instrument and mA outputs. All these values can be used to

evaluate the stability of the measurement and condition of the electronics and optical components of the

instrument.


Figure 4

Diagnostics tab shows the current diagnostic values and optical image

Optical Image and Slope

The optical image can be described as a window to the process medium. However, it also gives information

about the condition of the prism and the optics. Scratches or coating layers on the prism can be identified

through the optical image. Altering shapes are formed with every change in the prism-process interface.

Generally, the instrument has two different optical images, i.e. the raw image and IDS scaled optical image.

The raw image is the real response signal from the fotocells of the CCD-element, which is also unique to

every instrument. Consequently, the IDS image was developed to unify the optical images making the

instrument more user friendly. The IDS image is created by mathematical scaling of the raw image to a

rectangular shape improving the measurement stability on high and low refractive index (RI) ranges.