K-Patents PR-33-S Troubleshooting Guide User Manual

Page 8

Advertising
background image

TROUBLESHOOTING GUIDE
FOR PR-33-S AND PR-33-AC

8 (10)

January 9, 2015

K-PATENTS OY

Postal Address: P.O.Box 77, FI-01511 Vantaa, Finland

Street Address: Elannontie 5, FI-01510 Vantaa, Finland

Tel. int.+358 207 291 570

Fax int.+358 207 291 577

[email protected]

www.kpatents.com

VAT No. FI03035575

Business ID 0303557-5

Registered in Helsinki

Solution: Troubleshoot the issue indicated by the error message.

2. Description: No optical image, low image quality

Cause(s): The CCD-element does not receive enough light to determine the borderline of reflected light vs

shadow on the optical image. Lack of light in 90 % of the cases is caused by coated or eroded prism. Liquid

leakage through the prism gasket fits this category. Remaining 10 % of the faults are related to the other

optical or electrical faults, such as broken optical fiber or detached flat cable. Severely bad optical images

are generally seen with these error messages.

Solution: Clean the prism with a glass fiberbrush and check the optical image. If that does not help, check

the prism for any scratches or marks. The prism needs to be replaced if there are any visible marks on the

polished surfaces. The leakage can be detected from abnormal temperature readings and increased head

humidity. Services can be done only by K-Patents or a trained service person.

3. Description: Spiking concentration outputs

Cause(s): Coating of the prism or prism erosion causes a bad optical image and unstable measurement.

Solution: Clean the prism. If the prism is eroded, service accordingly. If the coating is a constant problem,

consider other installation position or additional washing system.

4. Description: Cannot receive signals from the sensor or establish connection through Ethernet

Cause(s): Faulty processor card, typically originates from weak cable connections or component failure due

to excessive humidity or temperature inside the sensor. Another potential cause may be a liquid leakage

inside the sensor.

Solution: The cables should be inspected, and if the problem is not resolved, the processor card needs to be

replaced by K-Patents or a trained service person. If the sensor has leaked liquid, most likely all the

electronics and optics need to be replaced.

5. Description: Inaccurate measurement

Cause(s): These issues are generally related to prism coating, which may result in a constantly increasing

concentration reading. It is also common that the fault is found in the reference measurement method (lab