TROUBLESHOOTING GUIDE
FOR PR-33-S AND PR-33-AC

5 (10)

January 9, 2015

K-PATENTS OY

Postal Address: P.O.Box 77, FI-01511 Vantaa, Finland

Street Address: Elannontie 5, FI-01510 Vantaa, Finland

Tel. int.+358 207 291 570

Fax int.+358 207 291 577

[email protected]

www.kpatents.com

VAT No. FI03035575

Business ID 0303557-5

Registered in Helsinki


Figure 5

IDS image on the left and raw image on the right

The detection of the measurement point is based on the steepest slope from the optical image (borderline of

activated vs inactivated light reflection). The slope image gives a derivative graph of the optical image that

can be used to examine the quality of the optical image and stability of the measurement.


Figure 6

Optical image and slope without sample

Without a sample on the prism surface the optical image should always return to “box” shape. Any outside

light can also affect the shape of the optical image, therefore, the prism should be covered when inspecting

the optical image. If the optical image is not shown as above, the prism should be cleaned with a glassfiber

brush, and inspected for scratches or marks.


Figure 7

Optical image and slope during measurement

During operation the optical image bends at a certain position which depends on the RI of the measured

medium. Also a clear dip appears on the slope image. A greater dip on the slope image demonstrates better

optical image quality and more stable measurement.