instrument).

Solution: Clean the prism of the sensor carefully and check the optical image. Proceed with the built-in

verification procedure. If the verification fails, the instrument needs to be inspected by qualified person

before recalibration.


Figure 9

Verification procedure for checking the calibration

Preventive maintenance inspection plan

Generally, the instrument is maintenance free, but in order to ensure a functioning K-Patents instrument, the

following preventive maintenance inspection plan is proposed. The weekly PMI plan includes only recording

of the diagnostic values without any service procedures. The annual plan includes also the verification

procedure and prism check for scaling or wear. In addition, there are some specific applications such as

aggressive solvents that require replacement of the flow cell o-ring gasket every 3 months.