TROUBLESHOOTING GUIDE
FOR PR-33-S AND PR-33-AC

6 (10)

January 9, 2015

K-PATENTS OY

Postal Address: P.O.Box 77, FI-01511 Vantaa, Finland

Street Address: Elannontie 5, FI-01510 Vantaa, Finland

Tel. int.+358 207 291 570

Fax int.+358 207 291 577

[email protected]

www.kpatents.com

VAT No. FI03035575

Business ID 0303557-5

Registered in Helsinki

The QF-value is related to the quality of the optical image and the slope (higher QF is better). However, the

sharpness of the slope and the QF-value can vary significantly depending on the application. The best way

to exploit the QF-values is to follow and record the changes in the QF which can reveal the changes on the

prism surface.

If the prism is damaged, there can be several dips on the slope which can disturb the measurement. As long

as the slope of the real measurement point is steep, the measurement is not lost. In the case of prism

coating the slope typically weakens which can lead to losing the measurement. However, low optical image,

high spikes, irregular form of the optical image or very weak slopes indicate that actions are required.

Typically, visual check of the prism condition and cleaning with a glassfiber brush is enough to resolve

measurement related problems. Replacement of the prism needs to be considered if there are any marks on

the prism surface.

Diagnostic Values

The instrument has several diagnostic values that can be used to inspect the condition of the instrument and

for troubleshooting purposes. Especially, the LED and the QF-value can be informative when interpreting the

optical images.

Explanation

Units / Range

CONC

Concentration

several units

TEMP

Temperature

°C / °F

CCD

Light activation of the CCD

%

nD

Refractive index

-

CALC

Calculated concentration

raw value

QF

Quality factor

typically 0-200

LED

Amount of light from the LED 0-100 %

HD TEMP Inside temperature

alarm at 65 °C

HD HUM

Inside humidity

alarm at 60 %

Table 1

Instrument diagnostic values

Interpretation of the Diagnostics

Measurement status: Indicates the current situation of the measurement. It also provides error messages.

The messages can be the following: Normal operation, No sample, Prism wash, Recovering, No sensor,

High sensor humidity, etc.. The complete list of different messages with descriptions can be found in the