OWON SDS Series User Manual

Page 26

Advertising
background image

5.Advanced User Guidebook

The description of the Channel Menu is shown as the following list:

Function
Menu

Setting

Description

Coupling

DC
AC
GROUND

Pass both AC and DC components of the input signal.
Block the DC component of the input signal.
Disconnect the input signal.

Inverted

ON
OFF

Display inverted waveform.
Display original waveform.

Probe

Attenu

X1
X10
X100
X1000

Match this to the probe attenuation factor to have an
accurate reading of vertical scale.

MeasCurr

YES
NO

If you are measuring current by probing the voltage
drop across a resistor, choose YES.

A/V (mA/V)
V/A (mV/A)

Turn the M knob to set the Amps/Volts ratio. The
range is 100mA/V ~ 1KA/V.
Amps/Volts ratio

= 1/Resistor value

Volts/Amp ratio is automatically calculated.

Limit

(SDS6062(V),

SDS7072(V)

does not have

this function)

full band
20M

Get full bandwidth.
Limit the channel bandwidth to 20MHz to reduce
display noise.

1. To set channel coupling

Taking the Channel 1 for example, the measured signal is a square wave signal
containing the direct current bias. The operation steps are shown as below:

(1) Press the CH1 MENU button to show CH1 SETUP menu.
(2) Press the H1 button, the Coupling menu will display at the screen.
(3) Press the F1 button to select the Coupling item as "DC". Both DC and AC

components of the signal are passed. The waveforms are shown as Figure
5-2
.

(4) Then, press F2 button to select the Coupling item as "AC". The direct current

component of the signal is blocked. The waveforms are shown as Figure 5-3.

21

Advertising
This manual is related to the following products: