successful unless it is within +/- 1 degree, however in the case of a field
calibration, this degree of accuracy may or may not be required. The degree
of accuracy that is acceptable is determined by the policy of the person
performing the calibration.

NOTE: Keep in mind that the DPC2530 only displays whole numbers, and
not tenths of a degree. Therefore, a reading of 18°F could be as low as 17.50
or as high as 18.49.


15.0

Allow the sensor to achieve equilibrium at ambient atmosphere.

15.1

After the 11.3% (Zero) calibration has been completed, remove the sensor

from the calibration salt and replace the cap.

15.2

Leave the sensor probe in the gland and while the unit is still on, allow it

to achieve equilibrium at the ambient atmosphere in the room. This should
take between two and three minutes. You will know when this has been
accomplished when the numbers on the display begin to stabilize.


16.0

Re-assemble the unit.

16.1

After the calibration process has been completed, remove the sensor probe

from the gland and return it to the sensor-sampling chamber, taking care to
position it properly. The white mark on the sensor probe should face towards
the right of the sensor-sampling chamber (at 3:00 if it were the face of a
clock.). If the white mark is not visible, then it should be placed so the sample
flow directly strikes the face of the mirror on the sensor tip (the sample flows
from right-to-left). In other words, the mirror should face the incoming gas
stream.


16.2

Hand-tighten the black sensor gland to prevent air from leaking out of the

sampling chamber.

16.3

Verify that the system is leak proof by turning on the pump and placing a

finger over the sample inlet port. The flow meter will drop to zero if there are
no leaks. If a leak is detected, make sure that all tubing connections are tight,
especially the black sensor gland.

16.4

After the unit has passed the leak test, the enclosure door can be closed.