The analyzer must be stored at ambient temperature (65-80°F) for at least four hours

prior to operation.

An in-line dryer for sample conditioning should be used for exothermic and

combustion applications.

This unit is not designed to measure the dew points in corrosive gasses, such as

Ammonia, S0

3

, Chlorine, and HCL.

Please read and understand this Product Manual before operating the unit.


Failure to comply with these conditions may cause damage to the unit that will not be
covered under the warranty. Super Systems, Inc. is not responsible for damage to this
unit caused by disregard of these warnings, neglect, or misuse.

TERMINAL BLOCK WIRING –

Power and communications wiring for the DPC2530 should be performed according to
the following diagram:

INSTRUMENT SETUP –


The instrument should be fully configured at the factory for immediate use. In the event
that the settings are ever lost, the following chart shows the appropriate input parameters
for the 7EK controller. When the instrument is set up in Fahrenheit using these
parameters, the input is scaled from –50 to +80°F and the 4-20mA output is scaled from –

FU100 - 2.0 AMP

FU103 - 1.0 AMP

1000

Positive

1002

Negative

GND

Ground

1101

Shield

1111

Positive

1121

Negative

1151

Positive

1161

Negative

120V 1 PHASE 60 HZ

SUPPLY POWER

4-20 mA

OUTPUT

RS 485 (MODBUS)

COMMUNICATIONS