Power supply and TE Device wire leads should be kept as short as possible to minimize electrical 
losses and reduce the likelihood of generating unwanted electromagnetic interference. Wire length 
must not exceed one meter. Use wires of a sufficient gage appropriate to the amount of electrical 
energy  each  wire  is  to  carry.  Wire  insulation  and  size  must  also  be  appropriate  to  the  ambient 
temperature  and/or  temperature  of  objects  in  contact  with  the  wire.  Wire  leads  supplied  by  TE 
Technology are for prototyping purposes and should be reviewed for appropriateness in the final 
application. 

Use  protection  devices  to  prevent  hazardous  conditions  and/or  damage  to  the  load  (e.g.  cooling 
assembly,  heater,  etc.)  and  other  related  equipment.  Protection  devices  must  operate 
independently of the temperature controller circuitry.  Protection devices should be placed at all 
points  on  the  load  and  related  equipment  where  a  hazardous  condition  can  be  detected.    These 
protection  devices  should  de‐energize  the  TC‐24‐10,  the  load,  and,  as  necessary,  other  related 
secondary equipment.  It is further recommended that such devices require the user to remove and 
correct the root cause of a fault before allowing the TC‐24‐10, the load, and related equipment to 
be re‐energized.  Protection devices should include, but are not limited to:  

Fuses to prevent against electrical overloads, 

Over/under  temperature  thermostats  to  prevent  against  hazardous  and/or  damaging 
temperatures, 

Liquid flow meters to prevent against damage due to loss of coolant flow 

 
The  TC‐24‐10 controller  (in  conjunction  with  the  standard  and  optional  sensors)  can not  detect 
under‐temperature  and  over‐temperature  conditions.  Therefore,  hazards  and/or  risk  of  loss  or 
damage  to  the  load  (e.g.  cooling  assembly,  heater,  etc.),  and/or  secondary  equipment  could 
occur if the temperature controller and/or sensors were to malfunction. Therefore, independent, 
redundant  protection  devices  are  recommended  in  addition  to  the  safeguards  provided  by  the 
temperature controller.   
 
 