TE Technology’s standard thermoelectric cooling assemblies (TCA) have at least one fan on the heat 
sink. The standard configuration has the thermoelectric modules and fan(s) wired to a terminal block 
with jumpers across the terminals so that the fans and modules are connected in parallel. However, 
this configuration is applicable only when applying power directly from the power supply. When using 
the  TCA  with  the  temperature  controller,  the  jumpers  MUST  be  removed  so  that  the  controller  is 
controlling  power  only  to  the  thermoelectric  modules.  There  must  be  no  electrical  connection 
between the fans and the modules; fans must be connected directly to the power supply, not to the 
controller.  The  controller  will  be  damaged  if  this  is  not  followed.  See  the  cooler  manual  for  further 
details, but the picture below shows the basic setup. 

 

 

 

 

Temperature Controller Connection

PROTECTION

DEVICES

DC POW ER SUPPLY

THERMOELECTRIC

(PELTIER)

HOT-SIDE FAN

(ALL UNITS)

COLD-SIDE FAN

(AC-XXX UNITS ONLY)

TCA

TEMPERATURE

CONTROLLER

Terminal Block

Cover removed

for clarity. Re-

install before

operation.

Review electrical jumper connections. Remove/install electrical jumpers as required.