Instruction Bulletin

106-300NE Rev. 3.4

May 2000

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-7

World Class 3000

VEE

DEFLECTOR

VEE

DEFLECTOR

DIFFUSION

ELEMENT

SETSCREW

FILTER

GAS FLOW

DIRECTION

APEX

P0017

Figure 2-2. Orienting the Optional Vee Deflector

c. Service

Required

1. Power input: 100, 115 or 220 Vac sin-

gle phase, 50 to 60 Hz, 3 amp
minimum. (See label.)

2. Compressed air: 10 psig (68.95 kPag)

minimum, 225 psig (1551.38 kPag)
maximum at 2 scfh (56.6 L/hr) maxi-
mum; supplied by one of the following
(less than 40 parts-per-million total hy-
drocarbons). Regulator outlet pressure
should be set at 5 psi (35 kPa).

(a) Instrument air - clean, dry.

(b) Bottled standard air with step-down

regulator.

(c) Bottled compressed gas mixture

(20.95% oxygen in nitrogen).

(d) Other equivalent clean, dry, oil-free

air supply.

Figure 2-3. Outline of Heater Power Supply