Probe troubleshooting, World class 3000 – Emerson WORLD CLASS 3000 User Manual

Page 46

Advertising
background image

Instruction Manual

Appendix A Rev. 3.6
July 1998

A-12 Appendices

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

PROBE TROUBLESHOOTING

A-7 OVERVIEW

The probe troubleshooting section describes
how to identify and isolate faults which may de-
velop in the probe assembly.

Install all protective equipment and
safety ground leads after trouble-
shooting. Failure to replace covers
and ground leads could result in seri-
ous injury or death.

A-8

PROBE TROUBLESHOOTING

a. Probe

Faults

Listed below are the four symptoms of
probe failure.

1. The system does not respond to

changes in the oxygen concentration.

2. The system responds to oxygen

changes but does not give the correct
indication.

3. The system does not give an accept-

able indication of the value of the oxy-
gen calibration gas being applied
during calibration.

4. The system takes a long time to return

to the flue gas value after the calibra-
tion gas is turned off.

b. Table A-2 provides a guide to fault finding

for the above symptoms.

c. Figure A-14 and Figure A-15 provide an al-

ternate approach to finding probe related
problems.

Table A-2. Fault Finding

Symptom

Check

Fault

Remedy

1. No response to oxy-

gen concentration
change when:

Thermocouple continuity

Thermocouple failure

Replace thermocouple
or return probe to
Rosemount.

Heater cold resistance to
be 11 ohm to 14 ohm

Heater failure

Replace heater or return
probe to Rosemount.

Heater is cold and TC
mV output is less than
set point

Triac O/P to heater

Failure of electronics

Check HPS and electron-
ics package.

Heater is hot and T/C
mV output is at set
point ±0.2 mV

Recorder chart

Recorder failure

See Recorder Instruction
Manual.

Cell mV input to electron-
ics and cell mV at probe
junction box

No cell mV at probe when
calibration gas applied

Replace cell or return
probe to Rosemount.

Probe cell mV OK but no
input to electronics

Check out cable
connection.

Cell mV satisfactory both
at probe junction box and
input to electronics - fail-
ure of electronics

Check electronics
package.

Advertising