Appendices A-17

World Class 3000

CELL EXTENSION

WIRE (ORANGE)

BOMB TAIL

CONNECTOR

INCONEL

CELL WIRE

(CLEAR

SLEEVING)

HEATER

WIRES

(BLACK

SLEEVING)

THERMOCOUPLE -

(RED ALUMEL)

THERMOCOUPLE +

(YELLOW CHROMEL)

CALIBRATION

GAS FITTING

REFERENCE

AIR FITTING

PROBE JUNCTION

BOX COVER

CABLE

27270019

Figure A-16. Cell Wiring Connection

b. If the probe uses the standard diffusion

element, use a spanner wrench to remove
the diffusion element.

c. If equipped with the optional ceramic diffusor

assembly, remove and discard setscrews,
Figure A-17, and remove vee deflector. Use
spanner wrenches from probe disassembly
kit, Table A-3, to turn hub free from retainer.
Inspect diffusion element. If damaged, re-
place element.

d. Loosen four socket head cap screws from

the cell and flange assembly and remove
the assembly and the corrugated seal. The
cell flange has a notch which may be used

to gently pry the flange away from the
probe. Note that the contact pad inside the
probe will sometimes fuse to the oxygen
sensing cell. If the cell is fused to the con-
tact pad, push the cell assembly back into
the probe (against spring pressure), and
quickly twist the cell assembly. The cell and
contact pad should separate. If the contact
pad stays fused
to the cell, a new con-
tact/thermocouple assembly must be in-
stalled. Disconnect the cell and the
thermocouple wires at the probe junction
box, and withdraw the cell with the wires still
attached (see paragraph A-13).