World class 3000 – Emerson WORLD CLASS 3000 User Manual

Page 47

Advertising
background image

Instruction Manual

Appendix A Rev. 3.6

July 1998

Rosemount Analytical Inc. A Division of Emerson Process Management

Appendices A-13

World Class 3000

Table A-2. Fault Finding (Continued)

Symptom

Check

Fault

Remedy

2. System responds to

oxygen concentration
changes but does not
give correct indication

Recorder or remote
indicator

Calibration error

Recalibrate recorder or
indicator. Reference Re-
corder Instruction Manual.

System calibration

Calibration error

Recalibrate system.

Probe mounting and con-
dition of duct

Air ingress into duct

Stop air leaks or resite
probe.

Good response, with
incorrect indication

Cell mV input to
electronics

Failure of electronics

Check electronics
package.

Calibration gas input port

Blocked port

Clean port. If the flue gas
is condensing in the cali-
bration gas line, insulate
the back of the probe.
Make sure that the cali-
bration gas line is capped
between calibrations, or a
check valve is installed.

3. System does not give

accurate indication of
applied calibration gas

Ceramic diffusion element

Diffusion element cracked,
broken, or missing

Replace diffusion element.

4. System takes a long

time to return to flue
gas value after calibra-
tion gas is turned off

Diffusion element

Plugged diffusion element

Change diffusion element
or snubber diffusion
element.