Miscellaneous, Maintenance / cleaning, Cabinet – Precision Medical PM4351 EasyFlow5 User Manual

Page 10: Warning

Advertising
background image

8

Miscellaneous

506639

– Nylon Hose Clamp

(Flowmeter & O2 Sensor)

505932

– 11/16” Pinch Clamp

(Sieve Bed Hose)

506763

– Tubing Sleeve Clamp

(Pressure Sensor)

507320 - #8 .625LG Hex Screw

(Sieve Beds to Base) 2 ea

505470 - #10 .750LG Hex Screw

(Cover to Base) 4 ea

506807 - Compressor Spring Pad

(Under Compressor each Spring) 2 per

506627 - #10 Phillips Screw

(Compressor Tray, Compressor Cover & Capacitor) 12 ea

506923 - Rubber Washer

(Under Sieve Beds) 2 ea

Maintenance / Cleaning

WARNING

Prior to cleaning, ensure the EasyFlow 5 Oxygen Concentrator is turned off and unplug.

DO NOT spray or apply any cleaners directly onto the device.

DO NOT allow any liquids to drip inside any case openings.

DO NOT use harsh and/or flammable chemicals to clean the device.

DO NOT clean with aromatic hydrocarbons.

Perform cleaning as required.

Cabinet:

1. Turn the EasyFlow 5 Oxygen Concentrator

to the “OFF” (O) position.

2. Unplug the device from the AC power source before cleaning.

3. Clean exterior surfaces of the device with a cloth dampened with disinfectant solution. Example: Sporicidin ®

Brand Disinfectant Solution.

4. Allow device to thoroughly air dry.

5. When not in use, store the device in a clean area free from grease, oil and other sources of contamination.