ETA Systems FC01-LQ User Manual

Page 54

Advertising
background image

Flow Meter FC01-LQ

Flow Meter FC01-LQ

48

Technical data

48

Technical data

8 Technical data

8.1 Ambient conditions

rail-mounted

surface mounted front panel mounted

Storage temperature:

-20 … +70 °C

-20 … +70 °C

-20 … +70 °C

Ambient temperature: *

+10 … +50 °C **

+10 … +50 °C

+10 … +50 °C

Degree of protection:

IP20

IP65

IP65

*

Only if the modules are spaced by at least 10 mm.

**

The max. ambient temperature of +40 °C applies to all systems fitted with current output C1.

8.2 Electrical characteristics

8.2.1 Power supply

DC supply

Pin selection:

Signal name

Pin XV

shield

1

+U

V

2

-U

V

3

8.2.1.1 DC voltage supply

Supply voltage:

U

VN

= DC 24 V *

Input voltage range:

U

V

= DC 19 V to DC 32 V

(ripple incl.)

Admissible ripple:

max. 20 % U

V

Rated current consumption:

I

vnk

= 170 mA with zero flow

I

vnk

= 200 mA with max. flow (end of measuring range)

Power consumption may be up to 300 mA ±10 % when analogue output C1 is fitted.

* DC 12 V possible if the FC01-LQ is used whithout option C1 (analogue outputs).

8 Technical data

8.1 Ambient conditions

rail-mounted

surface mounted front panel mounted

Storage temperature:

-20 … +70 °C

-20 … +70 °C

-20 … +70 °C

Ambient temperature: *

+10 … +50 °C **

+10 … +50 °C

+10 … +50 °C

Degree of protection:

IP20

IP65

IP65

*

Only if the modules are spaced by at least 10 mm.

**

The max. ambient temperature of +40 °C applies to all systems fitted with current output C1.

8.2 Electrical characteristics

8.2.1 Power supply

DC supply

Pin selection:

Signal name

Pin XV

shield

1

+U

V

2

-U

V

3

8.2.1.1 DC voltage supply

Supply voltage:

U

VN

= DC 24 V *

Input voltage range:

U

V

= DC 19 V to DC 32 V

(ripple incl.)

Admissible ripple:

max. 20 % U

V

Rated current consumption:

I

vnk

= 170 mA with zero flow

I

vnk

= 200 mA with max. flow (end of measuring range)

Power consumption may be up to 300 mA ±10 % when analogue output C1 is fitted.

* DC 12 V possible if the FC01-LQ is used whithout option C1 (analogue outputs).

Advertising