Flow Meter FC01-LQ

Flow Meter FC01-LQ

56

Technical data

56

Technical data

8.6 Sensor interface

8.6.1 Electrical data of the terminal for calorimetric monitoring heads

Terminal

Mnemonics

Data

XSK1

R(HEIZ)-LO

Function: terminal for negative pole of heater element
Drain output of heating current control
Max. sink current: I

sink

= 88 mA

Dielectric strength: -0.5 V … +20 V DC

XSK2

R(HEIZ)-HI

Function: terminal for positive pole of heater element
Hi-potential of heater source
Output voltage range (load dependent)
U

a

= 21 V … 24 V DC

Max. current output: I

max

= 100 mA

Not short-circuit proof

XSK3

R(Tref)-HI

Function: terminal for positive RTD * pole for medium
temperature measurement
Input resistance: > 1 G

Ω

Dielectric strength: -17 V … +30 V DC

XSK4

R(Tref)-LO

Function: terminal for negative RTD * pole for medium
temperature measurement
Input resistance: > 1 G

Ω

Dielectric strength: -17 V … +30 V DC

XSK5

AGND

Function: analogue ground
Reference potential of exitation current source for
RTD * operation

XSK6

IS

Function: output of exitation current source for
RTD * operation
Exitation current: 1 mA ± 1%
Admissible load range: R

load

= 0 … 2 k

Dielectric strength: ±15 V DC

XSK7

SGND

Function: shield ground

XSK8

Terminals for sensor cable shielding

XSK9

R(Tdiff)-LO

Function: terminal for negative pole of the heated RTD *
Input resistance: > 1 G

Ω

Dielectric strength: -17 V … +30 V DC

XSK10

R(Tdiff)-HI

Function: terminal for positive pole of the heated RTD *
Input resistance: > 1 G

Ω

Dielectric strength: -17 V … +30 V DC

* RTD = Resistive Temperature Device

8.6 Sensor interface

8.6.1 Electrical data of the terminal for calorimetric monitoring heads

Terminal

Mnemonics

Data

XSK1

R(HEIZ)-LO

Function: terminal for negative pole of heater element
Drain output of heating current control
Max. sink current: I

sink

= 88 mA

Dielectric strength: -0.5 V … +20 V DC

XSK2

R(HEIZ)-HI

Function: terminal for positive pole of heater element
Hi-potential of heater source
Output voltage range (load dependent)
U

a

= 21 V … 24 V DC

Max. current output: I

max

= 100 mA

Not short-circuit proof

XSK3

R(Tref)-HI

Function: terminal for positive RTD * pole for medium
temperature measurement
Input resistance: > 1 G

Ω

Dielectric strength: -17 V … +30 V DC

XSK4

R(Tref)-LO

Function: terminal for negative RTD * pole for medium
temperature measurement
Input resistance: > 1 G

Ω

Dielectric strength: -17 V … +30 V DC

XSK5

AGND

Function: analogue ground
Reference potential of exitation current source for
RTD * operation

XSK6

IS

Function: output of exitation current source for
RTD * operation
Exitation current: 1 mA ± 1%
Admissible load range: R

load

= 0 … 2 k

Dielectric strength: ±15 V DC

XSK7

SGND

Function: shield ground

XSK8

Terminals for sensor cable shielding

XSK9

R(Tdiff)-LO

Function: terminal for negative pole of the heated RTD *
Input resistance: > 1 G

Ω

Dielectric strength: -17 V … +30 V DC

XSK10

R(Tdiff)-HI

Function: terminal for positive pole of the heated RTD *
Input resistance: > 1 G

Ω

Dielectric strength: -17 V … +30 V DC

* RTD = Resistive Temperature Device