12.8 Troubleshooting

12.8.1 Error Messages

See “Error Messages” on page 151 for information relating to error messages.

12.8.2 Common Problems

and Solutions

Erratic Display

Check electrical connections for continuity and the transducer’s and/or amplifier’s
wiring code from its Certificate of Calibration and Application Note.

Incorrect wiring may have shorted the excitation power supply. Use the correct
wiring diagram to connect the transducer or amplifier to the instrument.

Make sure that the load on the transducer is constant.

Check input voltage or current to the (+)Signal (“+SIG”) and (-)Signal (“-SIG”) pins
with a multi-meter.

Blank Display

Incorrect wiring may have shorted the ±15 VDC excitation power supply. Use the
correct wiring diagram to connect the transducer or amplifier to the instrument.

+OVLD or -OVLD on Display

Indicates that the voltage across the (+)Signal (“+SIG”) and (-)Signal (“-SIG”) pins
is overranging or underranging the amplifier circuit. Make certain all wires are
connected properly.

If you remove all load from the transducer and you still see this message, the
(+)Excitation (“+EXC”) or (-)Excitation (“-EXC”) pins may be shorted to the (+)Sig-
nal (“+SIG”) or (-)Signal (“-SIG”) pins.

If you remove all load from the transducer and you get a numeric reading, the
transducer may have a high zero offset. Use the channel’s SETUP menu and set

DIAGNOSTICS

->

DISPLAY ADC

to “ON”; in the RUN mode this will allow the [VALUE] but-

ton to display raw A/D readings as a percentage of its full-scale. If the raw A/D
readings display more than +/-10% when there is no load on the transducer, the
transducer has a high zero offset.

Analog Output Incorrect

Make certain of the type of Analog Output (voltage or current) that the channel is
equipped with; see “Identifying the Output Type” on page 121.

Use the

DIAGNOSTICS

->

DAC FULL SCALE

and

DIAGNOSTICS

->

DAC ZERO SCALE

menu items to

force the Analog Output to a known output. Then, adjust your readout device,
panel meter, PLC or data acquisition system to match.

Check the output selection jumpers; see “Output Selection” on page 121.

Auxiliary Function Pins Not
Operating

Make sure that the

AUX1 FUNCTION

or

AUX2 FUNCTION

SETUP menu items are set cor-

rectly; if they are set to

DISABLED

then they will not operate.

The Auxiliary Function (“AUX1” and “AUX2”) pins must be connected to pin 10,
not pin 8, to activate them.

Sensitivity to EMI/RFI

To obtain maximum immunity to electromagnetic or radio frequency interference,
make certain that the shields of the transducer cables are connected to the “cable
shield connection screw” on the rear panel of the instrument. See “Wiring” on
page 98, “Extern
al Arrangement of AC powered SC1000 and SC2000” on
page 27 an
d “External Arrangement of Model SC3004” on page 33.