SC Series Instruction Manual page 71

Strain-

Gage

Input

C

ha

nnel

10

10.5.6

CALIBRATE

Menu

Item

This menu item performs a calibration according to what was entered in the

CALI-

BRATION TYPE

and

CALIBRATION DATA

menu items.

If the

CALIBRATION TYPE

is Shunt Calibration...

The display will read

DOING SHUNT CAL

, and prompt you to

APPLY 00000 UNIT

(where

00000 and UNIT are previously entered as

ZERO SCALE VALUE

and

DISPLAY.

UNIT

S. When you have applied this load to the transducer, press [ENTER].

The display will then read

WORKING

and calibration will take place. Upon com-

pletion, the display will indicate

DONE

and the instrument will return to the RUN

mode.

If the

CALIBRATION TYPE

is mV/V Calibration...

The display will read

DOING MV/V CAL

, and prompt you to

APPLY 00000 UNIT

(where

00000 and UNIT are previously entered as

ZERO SCALE VALUE

and

DISPLAY.

UNITS

). When you have applied this load to the transducer, press [ENTER].

The display will then read

WORKING

and calibration will take place. Upon com-

pletion, the display will indicate

DONE

and the instrument will return to the RUN

mode.

If the

CALIBRATION TYPE

is 2-Point Known Load Calibration...

The display will read

DOING 2POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where 00000 and UNIT are previously entered as

KNOWN POINT 1/2

and

DISPLAY.

UNITS

). When you have applied this load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are

previously entered as

KNOWN POINT 2/2

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

Upon completion, the display will indicate

DONE

and the instrument will return to

the RUN mode.

If the

CALIBRATION TYPE

is 3-Point Known Load Calibration...

The display will read

DOING 3POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where 00000 and UNIT are previously entered as

KNOWN POINT 1/3

and

DISPLAY.

UNITS

). When you have applied this load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00500 UNIT

(where 00500 and UNIT are

previously entered as

KNOWN POINT 2/3

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are

previously entered as

KNOWN POINT 3/3

and

DISPLAY. UNITS

). When you have

NOTICE

Before performing a calibration, the transducer must be connected to the
instrument, the

CALIBRATION TYPE

must be selected (see “CALIBRATION TYPE

Menu Item” on page 67), and the

CALIBRATION DATA

must be entered (see “CALI-

BRATION DATA

Sub-Menu” on page 69).

NOTICE

For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.