Configuring a Custom Cable

The Configure Custom Cable function allows you to define custom test standards
for up to four custom cables. You can change the following parameters for your
custom configuration:

•

Baseline standard for cable definition

•

NVP (Nominal Velocity of Propagation)

•

Maximum cable length

•

Fault threshold for impedance anomalies

•

Active pair assignments for cable pairs

•

Resistance

•

Characteristic impedance test enabled or disabled

•

Attenuation test enabled or disabled

•

NEXT and NEXT@REMOTE tests enabled or disabled

•

RL test enabled or disabled

•

ACR and ACR@REMOTE tests enabled or disabled

•

PSNEXT and PSNEXT@REMOTE tests enabled or disabled (Model
DSP-2000)

Note

To determine the NVP for a custom cable configuration refer to the
previous section, “NVP Calibration.”

During the configuration process, you can abandon the changes you have made by
pressing

! Abort. To reset a parameter back to the default value, highlight

the parameter; then press

@ Set to Default. To reset all of the

parameters back to default values, highlight the test standard on Page 1 of 6 of the
configuration screens; then press

@ Set to Default.