DSP-100,DSP-2000 Users

Manual Supplement

1/99

1

Change #1

On page 6-1, under

Calibrating the Test Tool, add the following note after

the first paragraph:

Note

Allow a 1 minute warm-up time prior to performing self-calibration (see
Chapter 6). Self-calibration should be performed only after the test tool
and remote unit have warmed up to an ambient temperature between
10°C and 40°C (50

°

F and 104

°

F).

On page 8-11, replace the

Return Loss specification with:

Return Loss (RL)

Return Loss range: 0 to 30 dB

Accuracy: ± 3 dB over the following ranges:

0 to 18 dB from 5 to 10 MHz

0 to 15 dB from 10.1 to 20 MHz

0 to 10 dB from 20.1 to 100 MHz