DSP-100/2000

Users Manual

4-22

Note

To extend battery life, use the ac adapter/charger when monitoring
impulse noise for extended periods. While monitoring impulse noise,
the test tool’s automatic power-down function is disabled.

Note

Monitoring impulse noise without a remote connected may produce
unreliable test results.

1.

If you are using another main unit as the remote, turn the remote unit’s rotary
switch to SMART REMOTE. If you are using a smart remote unit, turn its
rotary switch to ON.

2.

Use a 2m patch cable of the correct impedance to connect the remote to the far
end of the cable link.

3.

On Model DSP-100, remove any cable connected to the test tool’s BNC
connector.

4.

Turn the rotary switch to MONITOR.

5.

Use

D to highlight Impulse Noise on RJ45.

6.

Use a 2m patch cable of the correct impedance to connect the test tool to the
near end of the cable link. On Model DSP-2000, connect to the CABLE TEST
jack.

7.

Press

T to start the noise test.