SECTION 12—BATTERIES

Part No. 1118395

49

300 and 400

SECTION 12—BATTERIES

ƽ WARNING

Before performing any maintenance adjustment or service, turn power OFF and

remove key from ignition.
NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery

terminal(s)/post(s) at the same time. An electrical short may occur and serious

personal injury or damage may occur.
The use of rubber gloves and chemical goggles and face shield is recommended

when working with batteries.
Always use a battery lifting strap when lifting a battery. It is the most convenient

method and assures that the battery acid will not spill. It also helps to prolong the

life of the battery.
DO NOT tip the batteries. Keep the batteries in an upright position.
Invacare strongly recommends that battery installation and battery replacement

always be done by a qualified technician.
All battery terminals caps (two [2] on the LEFT battery and two [2] on the RIGHT

battery) MUST be installed prior to use.
Do not remove fuse or mounting hardware from the POSITIVE (+) RED battery

cable/mounting screw.
On 400 only - Battery terminal configuration shown below MUST be used. Batteries

that have the reversed terminal configuration MUST NOT be used - otherwise

serious injury or damage may occur.

ࣼ CAUTION

When connecting the battery cables to the batteries, the battery cables MUST be

connected to the battery terminals/posts as shown in FIGURE 12.2 for 300 and

FIGURE 12.5 for 400 - otherwise damage to the battery may result.
For proper battery connection, batteries MUST use post style terminals with

mounting holes through the terminal.

NEGATIVE (-)

Battery Terminal

POSITIVE (+)

Battery

Terminal

NEGATIVE (-)

Battery

Terminal

DO NOT USE

POSITIVE (+)

Battery

Terminal