SECTION 12—BATTERIES

Part No. 1118395

55

300 and 400

CONNECTING/DISCONNECTING THE BATTERY

CABLES ON THE 400

ƽ WARNING

On 400 only - Battery terminal configuration shown below MUST be used. Batteries

that have the reversed terminal configuration MUST NOT be used - otherwise

serious injury or damage may occur.

NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery

terminal(s)/post(s) at the same time. An electrical short may occur and serious

personal injury or damage may occur.
The use of rubber gloves and chemical goggles and face shield is recommended

when working with batteries.
Do not remove fuse or mounting hardware from the POSITIVE (+) RED battery

cable/mounting screw.
NEVER defeat or bypass any fuse. ONLY replace with a fuse of the same rating.

NOTE: For this procedure, refer to FIGURE 12.4 and FIGURE 12.5.

CONNECTING

1. Install battery terminal cap(s) onto battery cable(s) as follows (FIGURE 12.4):

A. RED battery terminal cap onto RED battery cable.
B. GREY battery terminal cap onto BLACK battery cable.

FIGURE 12.4 - INSTALLING BATTERY TERMINAL CAPS

NEGATIVE (-)

Battery Terminal

POSITIVE (+)

Battery

Terminal

NEGATIVE (-)

Battery

Terminal

DO NOT USE

POSITIVE (+)

Battery

Terminal

NOTE: Only battery cable and terminal cap

shown. Both caps install in the same manner

Battery Cable

Battery Terminal Cap