IBM R/S Rheometer M/02-212 User Manual

Page 40

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 40

M/02-212

For every step, there are the following inputs:
Input Step Nr 1

Start[1/s]: 0010.00
End[1/s]: 0100.00
Nr of MP:

010

Start Value

End Value

Number of MP (measur. points)

(input of numbers: see Chapter 6.4)

Minimum- and maximum-values of start values and end values depend on selected measuring system
for shear rate (D[s-1]) and shear stress (Tau[Pa]). At input the Rheometer checks start and end values
(if these values lie within due range) and indicates error massage “Range-Error” if out of range:

Range-Error

Range-Error
Start-Value:
min:

0.90

max:

1032.80

Name of Value out of range

Allowed mimum-value

Allowed maximum-value

This message will be indicated, for example, at input of the start value D[s

-1

] with measuring system

CC25, which lies outside the range of 0.9 ... 1032.8 s

-1

.

However, in the case, start value, end value and number of measurement points were input correctly,
the request for input of step duration will be indicated:

Input Step Nr. 1
time [s]: 0060

Step duration

Number of currently editing step

Minimum step duration:

D [s-1]

tmin = number of measurement points * 4 s

n [rpm]

tmin = number of measurement points * 4 s

Tau [Pa]

tmin = number of measurement points * 1 s

M [‰]

tmin = number of measurement points * 1 s

Maximum step duration allowed is: 3600 s

The duration value is automatically checked for the correct range (dependent on number of meas.
points). At wrong input, the “Range-Error” message will be indicated, accompanied by allowed values.

Attention: The more time between two measuring points, the higher precision of determination
of physical parameters!

This procedure of input Start-, End-Value, Number of MP, time is repeated for each step.

Finally the Program must be named and saved:
Edit PROGRAM name:

TEST

Input name of program

(for input of alpha-numerical texts see Chapter 6.4)