A2 error messages, Range error, Printer error – IBM R/S Rheometer M/02-212 User Manual

Page 58

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 58

M/02-212

A2 Error messages


The R/S Rheometer is rather error resistive, when errors occur they are determined and the user is
informed on the LCD display.
The most frequent error massages are explained in this Chapter.

Range Error

Cause:

User’s error at input. The user has tried to input a value that is less than the allowed
minimum value or is more than the allowed maximum value.


Message:


Example of error input of a preset value





What to do:

Input the value again. Be sure that new value is inside of the allowed value range.


Printer Error

Cause:

The printer is selected as output device and is not ready for printing.
a) Printer cable is not connected to the Rheometer
b) Paper out, printer not on line, other printer errors


Message:


Example: printer not ready




What to do:

Check if the printer cable is connected to the Rheometer.
Check if paper is in the feed sheeder of the printer.
Check if the “Ready” LED lights on printer.
Remove the error cause at printer and push key at Rheometer. If you cannot find the
error cause, push key in order to break the action that is carrying out now.

Range error
End value:
min : 0.90
max : 1032.80

Error #1
Printer not ready!

<OK> cont. <ST> stop

Name of value which is out of range

Allowed minimum value

Allowed maximum value

Cancel action

Try again

OK

ST

Advertising