Viii. technical data – IBM R/S Rheometer M/02-212 User Manual

Page 52

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 52

M/02-212

VIII. Technical data


R/S Rheometer

Dimensions

480 mm x 300 mm x 290 mm

Weight

8 kg

Nominal operating voltage
Power consumption (average)
Power consumption (maximum)

+/- 15 V, 5 V
12 W
22 W

Ambience conditions
Temperature
in operation
out of operation
Relative humidity (not condensable)
in operation
out of operation



10° to 40°C
10° to 45°C

20% to 80%
10% to 90%

Accuracy

± 1.0 % of maximum range value

± 1 digit

Torque range
Mains operation


0.05 to 50 mNm

Torque resolution

0.01 mNm

Speed range

0.7 min

-1

to 800 min

-1

Angle resolution

0.8 mrad

Temperature range

Depending on temperature control device used

-20°C to +180°C

Shear rate range

Depending on measuring system used

0.9 s

-1

to 4 × 10

3

s

-1

Shear stress range

Depending on measuring system used

0.7 Pa to 3.4 × 10

4

Pa

Viscosity range

The given range is a standard value (not maximum value)

Depending on measuring system used

1 × 10

-3

Pas to 3 × 10

3

Pas



AC-Adaptor

Dimensions

160 mm x 85 mm x 35 mm

Weight

0.5 kg

Power supply
Mains

voltage

secondary voltages (Output)

output-current
output

power


from 100 up to 240 V AC
5 V, +/- 15 V DC
2 A, 0.9 / -0.2 A
20 W

Frequency range of mains voltage

50 bis 60 Hz

Environment conditions
Temperature
in operation
out of operation

Relative humidity (not condensable)
in operation
out of operation



+10°C to +40°C
+10°C to +45°C

20% to 80%
10% to 90%

Advertising