Table Cone/Plate measuring systems (DIN 53018)
(consists of measuring cone and measuring device ME-CP/PP)

Measuring system C25-1

C25-2

C50-1

C50-2

C75-1

C75-2

Shear rate range
[s

-1

]

0 ... 4,800

0 ... 2,400

0 ... 4,800

0 ... 2,400

0 ... 4,800

0 ... 2,400

Shear stress range
[Pa]

0 ... 12,223

0 ... 12,223 0 ... 1,528

0 ... 1,528

0 ... 452

0 ... 452

Viscosity range
[Pas]

0.25 ... 1,629 0.5 ... 3,259 0.03 ... 203 0.06 ... 407 0.009 ... 60 0.02 ... 120

Filling

volume

[ml]

0.08 0.3 0.7 1.5 2.0 3.9

Shear rate factor K

γ⋅

[min/s]

6.00 3.00 6.00 3.00 6.00 3.00

Shear stress factor
τ

[Pa]

12.223 12.223 1.528 1.528 0.4527 0.4527

Radius of measu-
ring cone R [mm]

12.5 12.5

25 25 37.5

37.5

Angle of measu-ring
cone

α [°]

1 2 1 2 1 2

Cone truncation
[

µm]

50 50 50 50 50 50


Table Plate/Plate measuring systems (DIN 53018)
(consists of measuring plate and measuring device ME-CP/PP)

The following data is valid for a measuring gap of 1 mm. For other measuring gaps please calculate
K

γ⋅

with formula:

K

γ⋅

=(2

π/60) * (R/H) R= radius, H=gap


Measuring system

P25

P50

P75

Shear rate range [s

-1

]

0 ... 1,047

0 ... 2,094

0 ... 3,141

Shear stress range [Pa]

0 ... 16,298

0 ... 2,037

0 ... 603

Viscosity range [Pas]

1.56 ... 9,960

0.097 ... 623

0.02 ... 122

Filling volume [ml]

0.5

2.0

4.5

Shear rate factor K

γ⋅

[min/s]

1.309

2.6175 3.927

Shear stress factor

τ

[Pa]

16.2975

2.037 0.6036

Radius of measuring plate R [mm]

12.5

25

37.5

Advertising