Table Double Gap cylinder measuring system according to DIN 53453
(consists of measuring bob and measuring cup)

Measuring system

DG DIN

Shear rate range[s

-1

]

0 ... 4,031

Shear stress range [Pa]

0 ... 67

Viscosity range [Pas]

0.001 ... 1.30

Filling volume [ml]

22.5

Shear rate factor K

γ⋅

[min/s]

5.039

Shear stress factor

τ

[Pa]

0.067

Inner radius of measuring bob R

2

[mm]

22.75

Outer radius of measuring bob R

3

[mm]

23.5

Inner radius of measuring cup R

1

[mm]

22.25

Outer radius of measuring cup R

4

[mm]

24

Length of measuring bob L [mm]

111

Ratio of radii

δ

= Error! = Error!

1.021

Resistance coefficient c

L

1

The given ranges are standard values (not maximum values).

L

R

R

1

2

R

3

R

4

Measuring geometry