Zero calibration error, Break: mommax – IBM R/S Rheometer M/02-212 User Manual

Page 59

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 59

M/02-212

Zero Calibration Error

Cause:

An impermissible high value was measured during carrying out of zero point calibration of
the Rheometer



Message:



What to do:

Push key.
Check if the measuring system has been really removed from the Rheometer.
Retry zero point calibration.
If the error is presented again, that means that there is a fault in the measuring instrument


BREAK: MOMMAX


Cause:

Maximum torque has been exceeded during carrying out of a measurement.
a) You have selected inappropriate measuring range load for the measuring medium.
b) Rotation bob distorted in the measuring cup.


Message:

What to do:


Push key.
Check if the measuring system is fixed at the Rheometer properly. If so, select for this
measurement smaller shear rates or speed or use a measuring system with higher shear
stress factor.



Error #3
Zero cal. error!
Please retry cal.
<OK> cont. <ST> stop

BREAK: torque max!
Single mode
total MP: 10
total time: 60

ST

OK