Chapter 2

Connecting Signals

© National Instruments Corporation

2-5

SCXI-1503 User Manual

3-Wire Resistive Sensor Configuration

If you are using a 3-wire resistive sensor, you can reduce the error voltage
by one-half over the 2-wire measurement by connecting the device as
shown in Figure 2-3. In this configuration, very little current flows through
R

L3

and therefore R

L2

is the only lead resistance that introduces an error into

the measurement. The resulting measurement error is:

Figure 2-3. 3-Wire Resistive Sensor Configuration

VE

R

L2

I

EX

=

+

SCXI-1503

Channel X

IEX+

AI+

AI–

IEX–

SCXI-1306

External Sensor

ON

CH X

I = 100 µA

R

L2

R

L1

R

L3

R

T