Chapter 4

Theory of Operation

SCXI-1503 User Manual

4-6

ni.com

changes in resistance, you must use special configurations that minimize
measured errors caused by lead-wire resistance.

RTD Measurement Errors

Because the RTD is a resistive device, you must pass a current through the
device and monitor the resulting voltage. However, any resistance in the
lead wires that connect the measurement system to the RTD adds error to
the readings. For example, consider a 2-wire RTD element connected to a
measurement system that also supplies a constant current, I

EX

, to excite the

RTD. As shown in Figure 4-2, the voltage drop across the lead resistances
(labeled R

L

) adds an error voltage to the measured voltage.

Figure 4-2. 2-Wire RTD Measurement

For example, a lead resistance of 0.3

Ω in each wire adds a 0.6 Ω error to

the resistance measurement. For a platinum RTD at 0 °C with

α = 0.00385,

the lead resistance equates to an error of approximately

Chapter 2,

Connecting Signals

, describes different ways of connecting

resistive devices to the SCXI system.

The Relationship Between Resistance and
Temperature in RTDs

Compared to other temperature-measurement devices, the output of an
RTD is relatively linear with respect to temperature. The temperature
coefficient, called alpha (

α), differs between RTD curves. Although

various manufacturers specify alpha differently, alpha is most commonly

R

T

R

L

R

L

I

EX

+

V

0

0.6

Ω

0.385

Ω/°C

-----------------------------

1.6

°C

=