Table a-1. rtd measurement accuracy, Rtd measurement accuracy amplifier characteristics – National Instruments Module SCXI-1503 User Manual

Page 59

Advertising
background image

Appendix A

Specifications

SCXI-1503 User Manual

A-2

ni.com

Gain = 100

Calibrated

1

................................±25

μV max

±10

μV typ

With autozero enabled

2

............±10

μV max

±5

μV typ

Gain error (relative to calibration reference)

Gain = 1 or 100

Calibrated

1

................................0.074% of reading max

0.02% of reading typ

RTD Measurement Accuracy

Amplifier Characteristics

Input coupling.........................................DC

Input impedance

Normal powered on .........................>1 G

Ω

Input bias current ....................................±

2.8 nA

1

Assumes 1,000 point average, 25

°C ±10 °C over one year.

2

Assumes 1,000 point average, ±1

°C of autozero temperature.

Table A-1. RTD Measurement Accuracy

Measured

Temperature

°C

100

Ω Max °C

100

Ω Typ °C

1000

Ω Max °C

1000

Ω Typ °C

–100 to 0

0.60

0.23

1.09

0.46

0 to 25

0.62

0.23

1.11

0.47

25 to 100

0.69

0.25

1.20

0.49

100 to 500

1.11

0.37

1.68

0.65

500 to 1200

2.06

0.65

2.81

1.04

Notes: The accuracies in this table reflect using the module in4-wire mode. They do not include errors from the RTD
including lead-wire errors when using 2- or 3-wire connection.

The accuracies assume auto-zero is enabled and the environmental conditions are 25

°C ±10 °C over a one year period.

These accuracies were computed using a standard RTD with a TCR of 3851.