Although the resistance-temperature curve is relatively linear, accurately
converting measured resistance to temperature requires curve fitting. The
following Callendar-Van Dusen equation is commonly used to approximate
the RTD curve:

where

R

T

is the resistance of the RTD at temperature T.

R

0

is the resistance of the RTD at 0

°C.

A, B, and C are the Callendar-Van Dusen coefficients shown in
Table 4-1.

T is the temperature in

°C.

Table 4-1 lists the RTD types and their corresponding coefficients.

Table 4-1. Platinum RTD Types

Standard

Temperature

Coefficient of

Resistance

(TCR, PPM)

Typical

R

0

Callendar-Van

Dusen Coefficient

IEC-751
DIN 43760
BS 1904
ASTM-E1137
EN-60751

3851

100

Ω

1000

Ω

A = 3.9083

× 10

–3

B = –5.775

× 10

–7

C = –4.183

× 10

–12

Low cost
vendor
compliant

1

3750

1000

Ω

A = 3.81

× 10

–3

B = –6.02

× 10

–7

C = –6.0

× 10

–12

JISC 1604

3916

100

Ω

A = 3.9739

× 10

–3

B = –5.870

× 10

–7

C = –4.4

× 10

–12

US Industrial
Standard D-100
American

3920

100

Ω

A = 3.9787

× 10

–3

B = –5.8686

× 10

–7

C = –4.167

× 10

–12

R

T

R

0

1 AT BT

2

C T 100

(

)

3

+

+

+

[

]

=