Physical, Figure 6. scc-co20 dimensions, Maximum working voltage – National Instruments Isolated Current Output Module SCC-CO20 User Manual

Page 11: Isolation voltage

Advertising
background image

© National Instruments Corporation

11

SCC-CO20 Isolated Current Output Module User Guide

Physical

Figure 6. SCC-CO20 Dimensions

Weight .................................................... 28 g (1.0 oz)

I/O connectors ........................................ One 20-pin right-angle male

connector
One 3-pin screw terminal

Field-wiring diameter............................. 28 to 16 AWG

Maximum Working Voltage

(Signal + common-mode) each input should remain within 42.4 V

peak

or

60 VDC of ground.

Maximum working voltage refers to the signal voltage plus the CMV.

Channel-to-earth (inputs) ....................... 42.4 V

peak

or 60 VDC,

Measurement Category I

Caution

This device is rated for Measurement Category I and is intended to carry signal

voltages no greater than 42.4 V

peak

or 60 VDC. Do not use this device for connection to

signals or for measurements within Categories II, III, or IV.

Isolation Voltage

Channel-to-earth isolation

Continuous ...................................... 60 VDC,

Measurement Category I

Withstand ........................................ 2300 V

rms

verified by a 5 s

dielectric withstand type test

1.87 cm

(0.74 in.)

7.93 cm

(3.12 in.)

3.55 cm

(1.4 in.)

Advertising