What you need to get started, Safety information – National Instruments PXI Terminal Block NI TB-2708 User Manual

Page 2

Advertising
background image

NI TB-2708 Installation Guide

2

ni.com

The nine SMB connectors on the TB-2708 allow you to easily connect
analog input and analog output signals to your National Instruments device.
The 14-pin I/O connector located on the bottom of the TB-2708 allows
access to digital signals.

Caution

The TB-2708 is not designed for input voltages greater than 42 V

pk

/60 VDC,

even if you install a voltage divider that reduces the voltage to within the input range of the
DAQ device. Input voltages greater than 42 V

pk

/60 VDC can damage the TB-2708, any

device connected to it, and the host computer. Overvoltage also can cause an electric shock
hazard for the operator. National Instruments is not liable for damage or injury resulting
from such misuse.

What You Need to Get Started

You need the following to set up and use your terminal block:

NI TB-2708 Terminal Block Assembly

NI TB-2708 Installation Guide

One of the following:

NI PXI-6115

NI PXI-6120

S Series User Manual

Number 1 Phillips screwdriver

Safety Information

Cautions

Do not operate the device in an explosive atmosphere or where there may be

flammable gases or fumes.

Do not operate damaged equipment. The safety protection features built into this device
can become impaired if the device becomes damaged in any way. If the device is damaged,
turn the device off and do not use it until service-trained personnel can check its safety.
If necessary, return the device to National Instruments for service and repair to ensure that
its safety is not compromised.

Do not operate this equipment in a manner that contradicts the information specified in this
document. Misuse of this equipment could result in a shock hazard.

Advertising